Agilent E1441A Amplitude Flatness

Table D-26. Performance Test Record for the Agilent E1441A (Page 6 of 7)

Agilent E1441A Amplitude Flatness

Function

Output

Minimum

Measured

Maximum

M.U.

TAR

 

 

 

 

 

 

 

Sine wave

3.0V @ 1 kHz

reference

 

reference

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 100 kHz

ref - 0.030V

 

ref + 0.030V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 500 kHz

ref - 0.045V

 

ref + 0.045V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 1 MHz

ref - 0.045V

 

ref + 0.045V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 3 MHz

ref - 0.060V

 

ref + 0.060V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 5 MHz

ref - 0.060V

 

ref + 0.060V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 7 MHz

ref - 0.060V

 

ref + 0.060V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 9 MHz

ref - 0.060V

 

ref + 0.060V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 11 MHz

ref - 0.060V

 

ref + 0.060V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 13 MHz

ref - 0.060V

 

ref + 0.060V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Sine wave

3.0V @ 15 MHz

ref - 0.060V

 

ref + 0.060V

NA

Pass

 

 

 

 

 

 

/Fail

 

 

 

 

 

 

 

Agilent E1441A AM Modulation

Function

Output

Minimum

Measured

Maximum

M.U.

TAR

 

 

 

 

 

 

 

 

 

Sine wave

1.0V @ 1 kHz

0.495

Vrms

 

0.505

Vrms

0.8e-3

6.25:1

Modulated

100 Hz, 0%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sine wave

1.0V @ 1 kHz

0.6039

Vrms

 

0.6161

Vrms

0.8e-3

7.63:1

Modulated

100 Hz, 100%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 