Agilent Technologies E1441A Option 001 Specifications Timebase Accuracy, Phase Offset, Trigger

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 139
Image 139
Option 001 Specifications Timebase Accuracy:

Option 001 Specifications Timebase Accuracy:

Setability:

< 0.01 ppm

Stability:

±1 ppm 0°C - 50°C

Aging:

< 1 ppm in first 30 days (continuous operation)

 

10–7

 

month (after first 30 days)

 

---------------

Faceplate Input: (Ext Ref In terminal)

Lock Range:

10 MHz ±50 Hz

Level:

-10 dBm to +15 dBm,

 

+25 dBm or 10Vpp absolute maximum input

Impedance:

50±2%, 42 Vpk isolation from earth

Lock Time:

< 2 seconds

Faceplate Output: (Ref Out terminal)

Frequency:

10 MHz

Level:

>1 Vpp squarewave into 50

Phase Offset:

 

Range:

±360 degrees

Resolution:

0.001 degrees

Accuracy:

25 ns

Trigger:

 

Level:

5 V "zero-going" pulse

Pulse Width:

> 2 s

Fanout:

Four (4) Agilent E1441As (or Agilent E33120As)

Appendix A

Agilent E1441A Specifications 139

Page 139
Image 139
Agilent Technologies E1441A Option 001 Specifications Timebase Accuracy, Faceplate Output: Ref Out terminal, Phase Offset