AC Amplitude Adjustment (High-Z)

AC Amplitude Adjustment (High-Z)

The function generator stores twenty-two calibration constants related to HIGH Z output, and sixteen calibration constants related to 50 output. The constants are calculated from the adjustment value entered. The calibration constants are stored following completion of SETUP 22 (HIGH Z output) and the calibration procedure may be aborted after that point. No calibration constants are stored if the procedures are aborted at any other setup.

1.Use a DMM to measure the function generator ACrms output voltage for SETUP 08. These adjustments use a HIGH Z output termination.

Table D-13. AC Amplitude (High Z) Adjustment Setups

 

Nominal Output

 

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Adjustment for:

 

 

 

 

08

1 kHz

5.5 V rms

2 dB Output Attenuator

 

 

 

 

09

1 kHz

4.4 V rms

4 dB Output Attenuator

 

 

 

 

10

1 kHz

3.5 V rms

6 dB Output Attenuator

 

 

 

 

11

1 kHz

2.8 V rms

8 dB Output Attenuator

 

 

 

 

12

1 kHz

2.2 V rms

10 dB Output Attenuator

 

 

 

 

13

1 kHz

1.7 V rms

12 dB Output Attenuator

 

 

 

 

14

1 kHz

1.4 V rms

14 dB Output Attenuator

 

 

 

 

15

1 kHz

1.1 V rms

16 dB Output Attenuator

 

 

 

 

 

16

1 kHz

0.88

V rms

18 dB Output Attenuator

 

 

 

 

 

17

1 kHz

0.70

V rms

20 dB Output Attenuator

 

 

 

 

 

18

1 kHz

0.55

V rms

22 dB Output Attenuator

 

 

 

 

 

19

1 kHz

0.44

V rms

24 dB Output Attenuator

 

 

 

 

 

20

1 kHz

0.35

V rms

26 dB Output Attenuator

 

 

 

 

 

21

1 kHz

0.28

V rms

28 dB Output Attenuator

 

 

 

 

 

22

1 kHz

0.22

V rms

30 dB Output Attenuator

 

 

 

 

23

1 kHz

5.5 V rms

2 dB Pre-attenuator

 

 

 

 

24

1 kHz

4.4 V rms

4 dB Pre-attenuator

 

 

 

 

25

1 kHz

3.5 V rms

6 dB Pre-attenuator

 

 

 

 

26

1 kHz

2.8 V rms

8 dB Pre-attenuator

 

 

 

 

27

1 kHz

2.2V rms

10 dB Pre-attenuator

 

 

 

 

28

1 kHz

1.7 V rms

12 dB Pre-attenuator

 

 

 

 

 

29

1 kHz

1.4

Vrms

14 dB Pre-attenuator

 

 

 

 

 

2.Send the CAL:VALue <measured Vac> to the function generator.

3.Send CAL? command to the function generator to initiate calibration.

4.Repeat steps 1, 2 and 3 for SETUP 09 through 29.

5.Perform the “AC Amplitude Verification” on page 169.