Firmware Revision Query

Use the following command to perform a self-test:

*TST?

Returns “0” if the self-test passes or “1” if it fails. If the self-test fails, an error message is also generated with additional information on why the test failed. Use the SYSTem:ERRor? command to read the error queue.

Memory Locations Four memory locations (numbered 0, 1, 2 and 3) are available to store instrument configuration states. See the *SAV, *RCL and MEMory commands for more information.

Firmware Revision Query

SCPI Language Version Query

The function generator has two microprocessors for control of various internal systems. You can query the function generator to determine which revision of firmware is installed for each microprocessor.

The function generator returns two numbers. The first number is the firmware revision number for the main generator processor and the second is for the input/output processor.

Use the following command to query the firmware version:

*IDN?

Returns GPIB,E1441A,0,A.01.00-A.01.09

Be sure to dimension a string variable with at least 41 characters.

You can determine the SCPI version with which the function generator is in compliance by sending a query from the remote interface.

Use the following command to query the SCPI version:

SYSTem:VERSion?

Returns a string in the form “YYYY.V” where the “Y's” represent the year of the version, and the “V” represents a version number for that year

(for example, 1993.0).

56 Agilent E1441A Application Information

Chapter 2

Page 56
Image 56
Agilent Technologies E1441A user service Firmware Revision Query, SCPI Language Version Query