Function Gain and Linearity Verification

Function Gain and Linearity Verification

This test verifies the output amplitude accuracy specification for sine wave, triangle wave, ramp, and square wave outputs.

Set the function generator for each output indicated in the table below. Use a DMM to measure the function generator ACrms output voltage. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.

Table D-3. Function Gain and Linearity Verification

 

 

Agilent E1441A

 

Measurement

 

 

 

 

 

 

 

 

 

Function

 

Out

Ampl

Freq

Nominal

Error

 

 

 

Terma

 

 

 

 

 

 

 

 

 

 

 

 

Q

Sine wave

 

HIGH Z

7.0 Vrms

1.0 kHz

7.0 Vrms

± 0.07 Vrms

 

 

 

 

 

 

 

 

 

Sine wave

 

HIGH Z

5.7 Vrms

1.0 kHz

5.7 Vrms

± 0.057 Vrms

 

 

 

 

 

 

 

 

 

Triangle wave

 

HIGH Z

5.7 Vrms

100 Hz

5.7 Vrms

± 0.057 Vrms

 

 

 

 

 

 

 

 

 

Ramp wave

 

HIGH Z

5.7 Vrms

100 Hz

5.7 Vrms

± 0.057 Vrms

 

 

 

 

 

 

 

 

Q

Square wave

 

HIGH Z

10.0 Vrms

100 Hz

10.0 Vrms

± 0.1 Vrms

 

 

 

 

 

 

 

 

 

Square wave

 

HIGH Z

8.0 Vrms

100 Hz

8.0 Vrms

± 0.08 Vrms

 

 

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50 Ω ± 0.1 load on output.

DC Function Offset Verification

This test verifies the DC offset and DC output specifications.

Set the function generator for each output indicated in the table below. Use a DMM to measure the function generator DCV output. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.

Table D-4. DC Function Offset Verification

 

Agilent E1441A

 

Measurement

 

 

 

 

 

 

 

Function

Out

Ampl

Nominal

Error

 

 

Terma

 

 

 

Q

DC Volts

HIGH Z

10.0 VDC

10.0 VDC

± 0.20 VDC

 

 

 

 

 

 

 

DC Volts

HIGH Z

-10.0 VDC

-10.0 VDC

± 0.20 VDC

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50 Ω ± 0.1 load on output.

168 Service Procedures

Appendix D

Page 168
Image 168
Agilent Technologies E1441A Function Gain and Linearity Verification, DC Function Offset Verification, Service Procedures