FREQuency?

Parameters

Parameter

Parameter

 

Minimum

Maximum

Default

Name

Type

Function

Frequency

Frequency

Units

 

 

 

 

 

 

frequency

numeric

Sine

100 Hz

15 MHz

Hz

 

 

 

 

 

 

 

 

Square

100 Hz

15 MHz

Hz

 

 

 

 

 

 

 

 

Built-In Arbs

100 Hz

5 MHz

Hz

 

 

 

 

 

 

 

 

Ramp

100 Hz

100 kHz

Hz

 

 

 

 

 

 

 

 

Triangle

100 Hz

100 kHz

Hz

 

 

 

 

 

 

 

 

 

 

 

 

Comments For arbitrary waveforms that you create and download to memory,

the maximum frequency depends on the number of points specified in the waveform. As shown below, the maximum output frequency decreases as you specify more points in the waveform. The five built-in arbitrary waveforms can be output at a maximum of 5 MHz.

Number of Arb Points

Minimum Frequency

Maximum Frequency

 

 

 

 

 

 

8 to 8,192 (8K)

100 Hz

5 MHz

 

 

 

8,193 to 12,787 (12K)

100 Hz

2.5 MHz

 

 

 

12,288 to 16,000

100 Hz

200 KHz

 

 

 

 

 

 

Possible Conflict with Function Change: The output frequency is automatically adjusted if you select a function whose maximum frequency is less than that of the currently active function. For example, if you output a

1 MHz sine wave and then change the function to triangle wave, the function generator will adjust the output to 100 kHz (the upper limit for triangle waves). A -221, “Settings conflict” error is generated and the frequency is adjusted.

FREQuency?

[SOURce:]FREQuency? [MINMAX] queries the frequency setting for the function currently active. Returns a value in hertz.

FREQuency:STARt

[SOURce:]FREQuency:STARt <frequency>MINMAX sets the start frequency. Select from 10 mHz to 15 MHz (100 kHz for triangle and ramp). The sweep is phase continuous over the full frequency range. The default is 100 Hz. MIN = 10 mHz. MAX

=15 MHz. The start frequency setting is stored in volatile memory.