Table D-6. AC Amplitude Verification (50 Ohms)

Install the 50 feedthrough load between the DMM and the function generator output. Set the function generator for each output indicated in the table on the next page. Use a DMM to measure the ACrms output voltage of the function generator. Compare the measured results to the test limits shown in the table. This is a 50 output termination test.

Table D-6. AC Amplitude Verification (50 Ohms)

 

 

Agilent E1441A

 

Measurement

 

 

 

 

 

 

 

 

Function

Out

Ampl

Freq

Nominal

Error

 

 

Terma

 

 

 

 

Q

Sine wave

50

3.5 Vrms

1.0000 kHz

3.5 Vrms

± 0.035 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

2.8 Vrms

1.0000 kHz

2.8 Vrms

± 0.028 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

2.2 Vrms

1.0000 kHz

2.2 Vrms

± 0.022 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

1.7 Vrms

1.0000 kHz

1.7 Vrms

± 0.017 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

1.4Vrms

1.0000 kHz

1.4Vrms

± 0.014 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

1.1 Vrms

1.0000 kHz

1.1 Vrms

± 0.011 Vrms

 

 

 

 

 

 

 

Q

Sine wave

50

0.88 Vrms

1.0000 kHz

0.88 Vrms

± 0.0088 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.70 Vrms

1.0000 kHz

0.70 Vrms

± 0.0070 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.55 Vrms

1.0000 kHz

0.55 Vrms

± 0.0055 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.44 Vrms

1.0000 kHz

0.44 Vrms

± 0.0044 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.35 Vrms

1.0000 kHz

0.35 Vrms

± 0.0035 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.28 Vrms

1.0000 kHz

0.28 Vrms

± 0.0028 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.22 Vrms

1.0000 kHz

0.22 Vrms

± 0.0022 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.17 Vrms

1.0000 kHz

0.17 Vrms

± 0.0017 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.14 Vrms

1.0000 kHz

0.14 Vrms

± 0.0014 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.11 Vrms

1.0000 kHz

0.11 Vrms

± 0.0011 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.088Vrms

1.0000 kHz

0.088Vrms

± 0.00088 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.070 Vrms

1.0000 kHz

0.070 Vrms

± 0.00070 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.055 Vrms

1.0000 kHz

0.055 Vrms

± 0.00055 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.044 Vrms

1.0000 kHz

0.044 Vrms

± 0.00044 Vrms

 

 

 

 

 

 

 

Q

Sine wave

50

0.035 Vrms

1.0000 kHz

0.035 Vrms

± 0.00035 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.028 Vrms

1.0000 kHz

0.028 Vrms

± 0.00028 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.022 Vrms

1.0000 kHz

0.022 Vrms

± 0.00022 Vrms

 

 

 

 

 

 

 

 

Sine wave

50

0.018 Vrms

1.0000 kHz

0.018 Vrms

± 0.00018 Vrms

 

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50 Ω ± 0.1 load on output.

170 Service Procedures

Appendix D

Page 170
Image 170
Agilent Technologies E1441A user service Table D-6.AC Amplitude Verification 50 Ohms, Service Procedures, Appendix D