Agilent Technologies E1441A System Characteristics Configuration Times, Arb Download Times

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 138
Image 138
System Characteristics Configuration Times (2)

System Characteristics Configuration Times (2)

Function Change: (3)

 

80 ms

 

 

Frequency Change: (3)

10 ms

 

 

Amplitude Change: (3)

30 ms

 

 

Offset Change:

 

20 ms

 

 

Select User Arb:

 

550 ms

 

 

Modulation Parameter

 

 

 

 

Change:

 

<350 ms

 

 

Arb Download Times:

 

 

 

 

 

 

 

 

 

 

 

Table 3-1.

 

 

 

 

 

 

 

 

 

 

 

 

Arb Length

 

Binary

ASCII Integer

ASCII Real (4)

 

 

 

16,000 points

 

8 sec

81 sec

100 sec

 

 

 

 

 

 

 

 

 

 

 

8,192 points

 

4 sec

42 sec

51 sec

 

 

 

 

 

 

 

 

 

 

 

4,096 points

 

2.5 sec

21 sec

26 sec

 

 

 

 

 

 

 

 

 

 

 

2,048 points

 

1.5 sec

11 sec

13 sec

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(1)Trigger source ignored when External Gate is selected.

(2)Time to change parameter and output the new signal.

(3)Modulation or sweep off.

(4)Times for 5-digit and 12-digit numbers.

(5)See “Option 001 Specifications” on page 139.

General Specifications Agilent E1441A Available Power (Amps):

+5V: peak measurement (Ipm): 0.50 dynamic measurement (Idm): 0.10

+12V:peak measurement (Ipm):

2.50

dynamic measurement (Idm);

0.12

Cooling/Slot:

 

Average Watts/Slot:

25.0

deltaP mm H2O:

0.10

Air Flow liters/s:

2.00

Operating Environment:

0° C to 55° C (Option 001 to 50° C

 

65% Relative Humidity to 40° C

Storage Environment:

-40°C to 70° C

State Storage Memory: Four memory locations available (0, 1, 2 and 3)

Warm-up Time:

30 minutes

Programming Language:

SCPI-1993, IEEE-488.2

138 Agilent E1441A Specifications

Appendix A