Ext Trig / FSK / Burst Input Terminal

Ext Trig / FSK / Burst Input Terminal

command module by sending the IEEE-488 Group Execute Trigger (GET) message.

The input requirement for an external trigger, FSK input or Burst input is a TTL positive-going (when TRIG:SLOPE is POS) or negative-going (when TRIG:SLOPE is NEG) pulse that must be >2 microsecond in duration.

The Ext/FSK/Burst Input terminal is used in the following modes:

Figure 2-14.

Triggered Sweep Mode: Execute TRIG:SOUR EXT or TTLT<0-7> to enable the triggered sweep mode (sweeps must be enabled). When the edge (polarity set by TRIG:SLOPE) of a TTL pulse is received on the Ext Trig terminal or VXIbus TTLTRG line, the function generator outputs a single sweep.

Externally-Modulated FSK Mode: To enable the externally-modulated mode, execute FSK:SOUR EXT when FSK is enabled. When a low TTL level is present, the FSK carrier frequency is output. When a high TTL level is present, the FSK “hop” frequency is output.

Counted Burst Mode: Execute TRIG:SOUR EXT to enable the counted burst mode (burst must be enabled). When the rising edge of a TTL pulse is received, the function generator outputs a burst waveform with the specified number of cycles.

External Gated Burst Mode: To enable the external gated mode, execute BM:SOUR EXT when burst is enabled. When a high TTL level is present, the function generator outputs the carrier waveform. When a low TTL level is present, the output is disabled.