To Phase Lock Multiple

To Phase Lock Multiple

Function Generators

(requires Option 001)

The front-panel 10 MHz Ref Out and 10 MHz Ref In terminals allow you to synchronize multiple function generators. The following statements show you how to synchronize two function generators at 10 kHz (send the commands to both function generators):

APPL:SIN 10E+3, 5.0

Select sine function at 10 kHz

PHAS:ADJ

-90 DEG

Set phase offset to -90 degrees

PHAS:REF

 

Set phase reference to zero

 

 

 

 

 

 

 

 

 

 

Figure 2-12.

50 Agilent E1441A Application Information

Chapter 2

Page 50
Image 50
Agilent Technologies E1441A user service To Phase Lock Multiple Function Generators, requires Option, Figure, Chapter