Agilent Technologies E1441A user service DC Output Adjustment

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 184
Image 184
DC Output Adjustment

DC Output Adjustment

The function generator stores nine calibration constants related to DC volts output. The constants are calculated from the adjustment value entered. The calibration constants are stored following completion of SETUP 59. No calibration constants are stored if the procedures are aborted at any other setup.

1.Use a DMM to measure the function generator DCV output voltage for SETUP 50 in the following table. These adjustments use a HIGH Z output termination.

Table D-17. DC Output Adjustment Setups

 

Nominal Output

 

 

 

 

SETUP

DC Volts

Adjustment for:

 

 

 

50

- 8.0 VDC

Negative offset gain

 

 

 

51

8.0 VDC

Positive offset gain

 

 

 

52

0.0 VDC

AM offset

 

 

 

53

0.0 VDC

2 dB Pre-attenuator offset.

 

 

 

54

0.0 VDC

4 dB Pre-attenuator offset.

 

 

 

55

0.0 VDC

6 dB Pre-attenuator offset.

 

 

 

56

0.0 VDC

8 dB Pre-attenuator offset.

 

 

 

57

0.0 VDC

10 dB Pre-attenuator offset.

 

 

 

58

0.0 VDC

12 dB Pre-attenuator offset.

 

 

 

59

0.0 VDC

14 dB Pre-attenuator offset.

 

 

 

2.Send the CAL:VALue <measured Vdc> to the function generator.

3.Send the CAL? command to the function generator to initiate the calibration.

4.Repeat steps 1, 2 and 3 for SETUP 51 through 59.

5.Perform the “DC Function Offset Verification” on page 168.