Agilent Technologies E1441A FSKey:FREQuency?, FSKey:INTernal:RATE?, FSKey:SOURce, Chapter

Models: E1441A

1 206
Download 206 pages 30.53 Kb
Page 109
Image 109
FSKey:FREQuency

FSKey:FREQuency

[SOURce:]FSKey:FREQuency <frequency>MINMAX sets the FSK “hop” frequency. Select from 10 mHz to 15 MHz (100 kHz for triangle and ramp). The default is 100 Hz. MIN = 10 mHz. MAX = 15 MHz. The FSK frequency setting is stored in volatile memory.

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

frequency

numeric

10E-03 to 15E+06

Hz

 

 

 

 

 

 

 

 

FSKey:FREQuency?

[SOURce:]FSKey:FREQuency? [MINMAX] queries the FSK “hop” frequency. Returns a value in hertz.

FSKey:INTernal:RATE

[SOURce:]FSKey:INTernal:RATE <rate in Hz>MINMAX sets the rate at which the output frequency “shifts” between the carrier and hop frequency (FSK:SOUR INT only). Select from 10 mHz to 50 kHz. The default is 10 Hz. MIN = 10 mHz. MAX = 50 kHz. The FSK internal rate setting is stored in volatile memory.

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

rate

numeric

10E-03 to 50E+03

Hz

 

 

 

 

 

 

 

 

Comments The FSK rate is ignored when the external source is selected (FSK:SOUR EXT).

FSKey:INTernal:RATE?

[SOURce:]FSKey:INTernal:RATE? [MINMAX] queries the FSK rate. Returns a value in hertz.

FSKey:SOURce

[SOURce:]FSKey:SOURce INTernalEXTernal selects an internal or external FSK source. The default is INT. [ Stored in volatile memory ]

Comments When the internal source is selected, the rate at which the output frequency “shifts” between the carrier frequency and hop frequency is determined by the FSK rate specified (FSK:INT:RATE).

Chapter 3

Agilent E1441A SCPI Command Reference 109

Page 109
Image 109
Agilent Technologies E1441A user service FSKey:FREQuency?, FSKey:INTernal:RATE?, FSKey:SOURce, Parameters, Chapter