AM Modulation Depth Verification

AM Modulation Depth Verification

This test verifies the modulation depth specification.

Select each function generator output in the table below. Use a DMM to measure the function generator ACrms output voltage. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.

Table D-8. AM Modulation Verification

 

 

 

Agilent E1441A

 

 

 

Measurement

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AM Modulation

 

 

+

 

 

 

 

 

 

 

 

 

 

 

Function

Out

Ampl

Freq

Shape

Freq

Depth

Nominal

Error

 

 

Terma

 

 

 

 

 

 

 

Q

Sine wave

HIGH Z

1.0 Vrms

1.00 kHz

Sinewave

100 Hz

0%

0.50 Vrms

± 0.005 Vrms

 

 

 

 

 

 

 

 

 

 

 

Sine wave

HIGH Z

1.0 Vrms

1.00 kHz

Sinewave

100 Hz

100%

0.61 Vrms

± 0.0061 Vrms

 

 

 

 

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50 Ω ± 0.1 load on output.

Optional Performance Verification Tests

Square Wave Duty Cycle Verification

These tests are not intended to be performed with every calibration. They are provided as an aid for verifying additional instrument specifications.

This test verifies the duty cycle specification of the squarewave output.

Select each function generator output in the table below. Use an integrating DMM to measure the VDC output of the function generator. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.

NOTE: You make a DC voltage measurement on this test.

Table D-9. Square Wave Duty Cycle Verification

 

Function

Out

Ampl

Freq

Duty

Nominal

Error

 

 

Terma

 

 

Cycle

 

 

O

Square wave

HIGH Z

1.0 Vrms

300.00 Hz

50%

0.00 VDC

± 0.020 VDC

 

 

 

 

 

 

 

 

O

Square wave

HIGH Z

1.0 Vrms

300.00 Hz

25%

- 0.50 VDC

± 0.020 VDC

 

 

 

 

 

 

 

 

O

Square wave

HIGH Z

1.0 Vrms

300.00 Hz

75%

+ 0.50 VDC

± 0.020 VDC

 

 

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50 Ω ± 0.1 load on output.

Note

The DMM used for this test must be an integrating multimeter. If the first step does not measure 0 VDC, use an oscilloscope for this test.

Do not use an auto-ranging function on the DMM for this test. Fix the DMM measurement range at 10 VDC.

172 Service Procedures

Appendix D

Page 172
Image 172