Distortion Verification

Distortion Verification

This test checks the Harmonic Distortion at selected frequencies and harmonics. This test requires the use of a spectrum analyzer with dynamic range, frequency range, and resolution bandwidth adequate for the measurement.

Select each function generator output in the table below. Use a spectrum analyzer connected to the function generator output. Set the fundamental frequency reference to 0 dB and measure the 2nd through 5th harmonic frequencies relative to this reference. This test is a 50 output termination test.

Table D-10. Distortion Verification

 

 

 

 

 

 

 

Measurement

 

 

 

 

Agilent E1441A

 

 

 

 

 

 

 

 

 

 

 

 

harmonic

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Function

Out

Ampl

Freq

Fundamental

2nd

3rd

4th

5th

Amt below

 

 

Terma

 

 

 

 

 

 

 

reference

O

Sine wave

50

1.1 Vrms

20.00 kHz

reference

40 kHz

60 kHz

80 kHz

100 kHz

> 70 dB

 

 

 

 

 

 

 

 

 

 

 

O

Sine wave

50

1.1 Vrms

100.00 kHz

reference

200 kHz

300 kHz

400 kHz

500 kHz

> 60 dB

 

 

 

 

 

 

 

 

 

 

 

O

Sine wave

50

1.1 Vrms

1.00 MHz

reference

2 MHz

3 MHz

4 MHz

5 MHz

> 45 dB

 

 

 

 

 

 

 

 

 

 

 

O

Sine wave

50

1.1 Vrms

15.00 MHz

reference

30 MHz

45 MHz

60 MHz

75 MHz

> 35 dB

 

 

 

 

 

 

 

 

 

 

 

a.The E1441A has a fixed output impedance of 50 on the "Output" terminal. Use the OUTPut:LOAD 50 INFinity MIN MAX command to set the output termination. HIGH Z assumes no load on output. 50 assumes a 50

± 0.1 load on output.

Appendix D

Service Procedures 173

Page 173