in the waveform (see below).

Voffset

Vpp

V

 

and

+ -------

max

2

 

 

Voffset

2Vpp

You can substitute "MINimum", "MAXimum" or "DEFault" in place of a specific value for any of the above three parameters. MIN selects the lowest value, MAX selects the highest value and DEF selects the default setting.

Output Frequency Minimum frequency is the same for all functions. Maximum frequency is dependent

on the function selected as shown in the following table.

Function

Minimum

Maximum Frequency

 

Frequency

 

 

 

 

Sine

100 Hz

15 MHz

 

 

 

Square

100 Hz

15 MHz

 

 

 

Built-In Arbs a

100 Hz

200 kHz, 2.5 MHz or 5 MHz b

 

 

 

Ramp

100 Hz

100 kHz

 

 

 

Triangle

100 Hz

100 kHz

 

 

 

 

 

 

a.There are five built-in arbitrary waveforms stored in non-volatile memory: sinc, negative ramp, exponential rise, exponential fall and cardiac. These waveforms are shown on page 48. The five built-in arbitrary waveforms can be output at a maximum of 5 MHz.

b.Maximum frequency for arbitrary waveforms you generate depends on the number of points

specified in the waveform. The maximum output frequency decreases as you specify more points

in the waveform (see below).

.

Number of Arb Points

Minimum Frequency

Maximum Frequency

 

 

 

8 to 8,192 (8k)

100 Hz

5 MHz

 

 

 

8,193 to 12,287 (12k)

100 Hz

2.5 MHz

 

 

 

12,288 to 16,000

100 Hz

200 kHz

 

 

 

 

 

 

Possible Output Frequency Conflict with a Function Change: The output frequency is automatically adjusted if you change the function from SINE or SQUARE (MAX frequency = 15 MHz) to RAMP, TRIANGLE or a Built-In Arb, if the current frequency is greater than that of the maximum frequency of the new function. For example, the maximum frequency of a sine wave is 15 MHz while the maximum frequency for a triangle wave is 100 kHz. If the current output is a sine wave at 1 MHz and you change the function to triangle wave which cannot output at 1 MHz, the function generator automatically adjusts the output to the upper frequency limit for a triangle wave (100 kHz). This function change will also generate a -221 "Settings conflict" error.

Possible Conflict with Duty Cycle (square wave only): For output frequencies above 5 MHz, the duty cycle is limited to values between 40% and 60% (below 5 MHz, the range is 20% to 80%). The duty cycle is

Chapter 3

Agilent E1441A SCPI Command Reference 97

Page 97
Image 97
Agilent Technologies E1441A user service in the waveform see below