Table D-22. AC Flatness Adjustment Setups

3.Send the CAL:VALue <measured Vac> to the function generator.

4.Send the CAL? command to the function generator to initiate the calibration.

5.Configure SETUP 65 in Table D-22 and repeat steps 3 and 4 for SETUP 65 through 82 in table D-22.

Table D-22. AC Flatness Adjustment Setups

 

Nominal Output

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Adjustment for:

 

 

 

 

65

100 kHz

3.0 V rms

100 kHz amplitude flatness

 

 

 

 

66

500 kHz

3.0 V rms

500 kHz amplitude flatness

 

 

 

 

67

1 MHz

3.0 V rms

1 MHz amplitude flatness

 

 

 

 

68

3 MHz

3.0 V rms

3 MHz amplitude flatness

 

 

 

 

69

5 MHz

3.0 V rms

5 MHz amplitude flatness

 

 

 

 

70

7 MHz

3.0 V rms

7 MHz amplitude flatness

 

 

 

 

71

9 MHz

3.0 V rms

9 MHz amplitude flatness

 

 

 

 

72

10 MHz

3.0 V rms

10 MHz amplitude flatness

 

 

 

 

73

10.5 MHz

3.0 V rms

10.5 MHz amplitude flatness

 

 

 

 

74

11 MHz

3.0 V rms

11 MHz amplitude flatness

 

 

 

 

75

11.5 MHz

3.0 V rms

11.5 MHz amplitude flatness

 

 

 

 

76

12 MHz

3.0 V rms

12 MHz amplitude flatness

 

 

 

 

77

12.5 MHz

3.0 V rms

12.5 MHz amplitude flatness

 

 

 

 

78

13 MHz

3.0 V rms

13 MHz amplitude flatness

 

 

 

 

79

13.5 MHz

3.0 V rms

13.5 MHz amplitude flatness

 

 

 

 

80

14 MHz

3.0 V rms

14 MHz amplitude flatness

 

 

 

 

81

14.5 MHz

3.0 V rms

14.5 MHz amplitude flatness

 

 

 

 

82

15 MHz

3.0 V rms

15 MHz amplitude flatness

 

 

 

 

6. Perform the “Amplitude Flatness Verification” on page 171.

Appendix D

Service Procedures 187