Klark Teknik DN9340E, DN9344E manual To switch to the Parametric Equalisation view, User Guide

Models: DN9344E DN9340E

1 65
Download 65 pages 57.17 Kb
Page 16
Image 16

HELIX DN9340E/DN9344E Remote Control Software

User Guide

5.2Parametric Equalisation

To switch to the Parametric Equalisation view

Click on the

Para

button of the Channel A or B Navigation Strip, as required. (If staying on

the same Channel, you may prefer to click on a Parametric tab).

PEQ sections are adjusted by dragging the ‘target’ or setting values

PEQ panels appear as you add PEQ sections

Tick the check box to switch the

PEQ section out of the circuit

Parametric Mode tab

The black graph shows combined EQ. The red graph shows the response of the selected PEQ section

Parametric

Shaping tab

The upper Parametric Mode tab is used to set the EQ characteristic (EQ type), bypass state and graph display mode (when a DN9340) for the current channel. The lower Parametric Shaping tab provides a series of PEQ control panels for added PEQ sections. PEQ sections can also be adjusted at the display by using the ‘target cursor’.

To set the Parametric EQ mode for the channel

1Select a Proportional, Constant, or Symmetrical Q

characteristic from the EQ Type drop down menu (right).

2Use the Unit Display drop down menu to set the Parametric display mode that the HELIX unit should assume when taken out of slave operation; this only comes into effect if the slave is a DN9430. The choice made here has no effect on your AddIn device display.

Note Refer to the DN9340/DN9344 Operators Manual for application notes on Equalisation modes and Q.

16 of 65

Page 16
Image 16
Klark Teknik DN9340E To switch to the Parametric Equalisation view, To set the Parametric EQ mode for the channel