Sigma LBA-500PC, LBA-710, LBA-714PC, LBA-300, LBA-712 manual = ∑x ∑y y − y2 ⋅ Zx, y, Knife Edge Method

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 133
Image 133

The standard deviations are derived from the variances of the energy distributions and are equal to the standard deviations squared. The variances are:

σ2x

σ2y

=x y (x x )2 Z(x, y)

x y Z(x, y)

=x y ( y y)2 Z(x, y)

x y Z(x, y)

Where:

Z = the intensity of the pixel

x and y are the coordinates of the centroid

Only beam propagation factors based on beam widths and divergence angles derived from the second moments of the energy density distribution function, will allow one to predict how a beam will propagate. Other definitions of the beam widths and divergence angles may be used, but they must be shown to be equivalent to the second moment definitions for computing the correct beam propagation.

To make an accurate measurement of the beam widths with the LBA-PC you must aperture the beam inside a Drawn or Auto Aperture. The aperture must be approximately 2x the size of the beam. The Auto Aperture feature of the LBA-PC will automatically provide such an aperture under most operating conditions. It can be used in combination with a Drawn Aperture if needed. If your beam size is already equal to about 1/2 the beam display window, then you can usually get by without drawing an aperture, just be sure to center the beam in the window.

6.11.2Knife Edge Method

Beam widths are computed using special algorithms that simulate knife-edge techniques. The method employed in the LBA-PC borrows from two sources. They are:

ISO 11146 Lasers and laser-related equipment—Test methods for laser beam parameters— Beam Widths, divergence angle and beam propagation factor

Note this document is being revised and split into three parts:

11146-1, Part 1: Stigmatic and simple astigmatic beams

11146-2, Part 2: General astigmatic beams

11146-3, Part 3: Alternative test methods and geometrical laser beam classification and propagation

And from the IEEE Journal of Quantum Electronics, Vol 27, No 4, April 1991 Choice of Clip Levels for Beam Width Measurements Using Knife-Edge Techniques by Siegman, Sasnett and Johnston.

The LBA offers the operator two methods for computing Knife Edge beam widths. The 90/10 method presets the Clip% values to 90% and 10% respectively, and the Multiplier to 1.561. These are the recommended values based upon the above Siegman, et al. paper, and are very compatible with CCD camera noise figures. These values are perfectly correct for computing an

Operator’s Manual

133

LBA-PC

Doc. No. 10654-001, Rev 4.10

 

 

Page 133
Image 133
Sigma LBA-500PC, LBA-710, LBA-714PC, LBA-300, LBA-712, LBA-708, LBA-700, LBA-400 manual = ∑x ∑y y − y2 ⋅ Zx, y, Knife Edge Method