Sigma LBA-700, LBA-710, LBA-714PC, LBA-300 FRM and RDD are ignored if the LBA-PC is Not in remote

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 222
Image 222

 

Key

Type

Value Description

 

 

 

 

 

 

 

DataLogging

B

Data logging enable.

 

 

 

 

 

 

 

DataFileName

S

Name of data log file.

 

 

 

 

If no extension (i.e., no period) then

 

 

 

 

.LB3/4/5/7 is appended.

 

 

 

 

Maximum 256 characters.

 

 

 

 

If this key is set to “FRM” or “RDD” (case

 

 

 

 

ignored) (or was previously set to FRM or

 

 

 

 

RDD in the dialog) then data frames will be

 

 

 

 

logged to the GPIB bus. Each time a new

 

 

 

 

frame is captured the LBA-PC will

 

 

 

 

automatically send a response as if you had

 

 

 

 

sent “:FRM?” or “RDD?” respectively. You

 

 

 

 

must address the LBA-PC to talk and be

 

 

 

 

prepared to accept the transmitted data.

 

 

 

 

The LBA-PC will wait until you have read the

 

 

 

 

data before capturing another frame. Note:

 

 

 

 

FRM and RDD are ignored if the LBA-PC is

 

 

 

 

not in remote.

 

 

 

 

 

 

 

ResultsLogging

B

Results logging enable.

 

 

 

 

 

 

 

ResultsFileName

S

Name of results log file.

 

 

 

 

If no extension (i.e., no period) then .RLG is

 

 

 

 

appended.

 

 

 

 

Maximum 256 characters.

 

 

 

 

If this key is set to “RDR” (case ignored) (or

 

 

 

 

was previously set to RDR in the dialog) the

 

 

 

 

results will be logged to the GPIB bus. Each

 

 

 

 

time results are computed for a new frame

 

 

 

 

the LBA-PC will automatically send a

 

 

 

 

response as if you had sent “:RDR?

 

 

 

 

Values=1”. The LBA-PC will wait until you

 

 

 

 

have read the data before capturing another

 

 

 

 

frame. Note, in this case the results format

 

 

 

 

is ignored. Note: RDR is ignored if the LBA-

 

 

 

 

PC is not in remote.

 

 

 

 

 

 

 

ResultsFormat

L

Results log format.

 

 

 

 

0 = math

 

 

 

 

1 = spreadsheet

 

 

 

 

 

 

 

ExportLogging

B

Export logging enable.

 

 

 

 

 

 

 

ExportFileName

S

Name of export log file. Any file extension is

 

 

 

 

ignored and an appropriate extension is used

 

 

 

 

depending on the type of export file that is

 

 

 

 

being logged.

 

 

 

 

 

 

 

BMP

B

.BMP logging enable.

 

 

 

 

 

 

Operator’s Manual

 

222

LBA-PC

Page 222
Image 222