Sigma LBA-712, LBA-710, LBA-714PC, LBA-300, LBA-708 DSF display frame, ERR error reporting, Err?

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 243
Image 243

Key

Type

Value Description

 

 

 

X

F

x location in world coordinates5

 

 

 

Y

F

y location in world coordinates5

 

 

 

Z

F

Value of pixel at cursor

 

 

 

Delta

F

distance from cross hair to cursor

 

 

 

A.5.4.4 DSF - display frame

:DSF <configuration>3 :DSF?

Key

Type

Value Description

 

 

 

FrameNumber

I

frame number

 

 

-1 = gain frame

 

 

0 = reference frame

 

 

1 to n. ‘n’ is the size of the frame buffer

 

 

default = current frame

 

 

 

A.5.4.5 ERR - error reporting

:ERR <configuration>

Key

Type

Value Description

 

 

 

Verbose

B

true = remote error messages

 

 

if a communications error occurs then the

 

 

EMAV bit is set in the status byte (STB) and

 

 

a text message containing a brief

 

 

explanation of the error is placed in the

 

 

error message queue. To retrieve error

 

 

messages use “:ERR?”. All error messages

 

 

begin with the characters “!!!”.

 

 

default = true

 

 

 

:ERR?

A separate output queue is maintained for error messages. Each time :ERR? is received the LBA- PC returns the next message in the error message queue. If the queue is empty the LBA-PC returns the current verbose setting, i.e. “ERR Verbose=b”. The status of the error message

Operator’s Manual

243

LBA-PC

Doc. No. 10654-001, Rev 4.10

 

 

Page 243
Image 243
Sigma LBA-712, LBA-710, LBA-714PC, LBA-300, LBA-708, LBA-500PC, LBA-700, LBA-400 DSF display frame, ERR error reporting, Err?