Sigma LBA-500PC EXP set export configuration & export images, LOG set logging configuration, Bmp

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 221
Image 221

A.5.1.5 EXP - set export configuration & export image(s)

:EXP <configuration>1 :EXP?

Key

Type

Value Description

 

 

 

Export3

B

false = set configuration only.

 

 

true = set configuration then export images.

 

 

Images cannot be exported while running.

 

 

default = true

 

 

 

Replace

B

Overwrite if file exists.

 

 

default = false

 

 

 

ExportFileName6

S

Base name of export file. Extension

 

 

automatically created by LBA-PC based on

 

 

export file type.

 

 

Maximum 256 characters.

 

 

 

StartFrame

I

First frame to export.

 

 

1 to n. ‘n’ is the size of the frame buffer.

 

 

 

NumberFrames

I

Number of frames to export.

 

 

0 = all frames

 

 

1 to n. ‘n’ is the size of the frame buffer.

 

 

 

BMP

B

.BMP export enable.

 

 

 

AsciiComma

B

.CMA export enable.

 

 

 

AsciiSpace

B

.SPC export enable.

 

 

 

CursorData

B

.CUR export enable.

 

 

 

ColumnRowSum

B

.SUM export enable.

 

 

 

A.5.1.6 LOG - set logging configuration

NOTE: There is no default path for this command. If you want the data log file, results log file, or export log file(s) to be saved in a particular path then send the path with the file name.

Data & Results logging will append to an existing file once opened. It will overwrite an existing file if newly opened. Export logging will always overwrite an existing file. No warning or error message is given in either instance.

:LOG <configuration>1,3 :LOG?

Operator’s Manual

221

LBA-PC

Doc. No. 10654-001, Rev 4.10

 

 

Page 221
Image 221