Sigma LBA-500PC, LBA-710, LBA-714PC, LBA-300, LBA-712, LBA-708, LBA-700, LBA-400 manual LOG Logging

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 197
Image 197

Centroid X,

Centroid Y,

Width X,

Width Y,

Diameter(^END)

LBA-PC sends

RDR ,%,,,um,um,um,um,um,um,um(^END)

10.6.2.6 LOG - Logging

You can configure the LBA-PC to automatically send a FRM or RDD and RDR response each time new data is acquired. Remote logging is the fastest method available to transfer new frames from the LBA-PC.

Use the File Logging… dialog box or the LOG configuration command to specify logging new data or results to the host. If the data log file name is “FRM” or “RDD” then each time a new frame of data is acquired the LBA-PC will automatically send a response as if you had sent “:FRM?” or “:RDD?” respectively. If the results log file name is “RDR” then each time a new frame of data is acquired the LBA-PC will automatically send a response as if you had sent “:RDR?”.

The LBA-PC will not capture another frame until the host reads the current logging response. This example describes how to enable data file logging to the host controller.

Host sends

:LOG

DataLogging=1;

DataFileName=FRM(^END)

Host sends

:RUN

LBA-PC sends

FRM FrameNumber=1; #516384(DAB)…(DAB) (^END)

LBA-PC sends

FRM

FrameNumber=2;

Operator’s Manual

197

LBA-PC

Doc. No. 10654-001, Rev 4.10

 

 

Page 197
Image 197
Sigma LBA-500PC, LBA-710, LBA-714PC, LBA-300, LBA-712, LBA-708, LBA-700, LBA-400 manual LOG Logging