5.Click on the Advanced tab, make sure that Automatic Serial Polling is not checked.

6.Click OK.

NOTE: It is possible for the LBA-PC to generate many service requests per second and the NI-488.2M default is to queue service requests. For these reasons we suggest you disable Automatic Serial Polling on both the LBA-PC and the host controller.