Sigma LBA-712, LBA-710 Special Setup for Pyrocam I Operation, Pyrocam I with Non-Digital LBA-PC’s

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 27
Image 27

2.6 Special Setup for Pyrocam I Operation

You must use special setups if you want to successfully interface your Pyrocam I with a Model LBA-PC frame grabber system. It is strongly recommended that you first become familiar with the operating characteristics of both your Pyrocam I and LBA-PC before attempting to operate them together.

To operate your Pyrocam I with a model LBA-300PC or a LBA-400/500/708/710/712/714PC without a digital camera option, see section 2.12.1.

To operate your Pyrocam I with model LBA-500PC-D or a model LBA-7XXPC-D, with the digital camera option, see section 2.12.2.

2.6.1Pyrocam I with Non-Digital LBA-PC’s

LBA-300/400/500/708/710/712/714PC

w/o digital camera option

2.6.1.1Pyrocam I Setup Requirements:

2.6.1.1.1Configure Video Output

The Pyrocam must be configured to output monochrome video in CCIR format. To do this, Dip Switch number 6 must be in the ON position. See Chapter 6 and Appendix A-6 in your Pyrocam Operator’s Manual for details.

2.6.1.1.2Configure Gain and Vertical Scale

Do not operate the Pyrocam I with the Gain switch set in the 10 position or with the Vertical Scale set in the Auto or Manual x8 positions. Use only the 1 or 3 Gain setting and Manual x1, x2 or x4 settings. The LBA will not be able to Ultracal in the gain equals 10 or scale x8 setting positions, and the Ultracal will not remain valid if the Pyrocam is in the Auto scale mode.

Note: You will need to temporarily attach your Pyrocam to either a VGA or a monochrome monitor to observe the Vertical Scale settings.

2.6.1.1.3Configure MONO/DIG/VGA switch to MONO position.

2.6.1.1.4Install Video Cable

Connect the Pyrocam’s Video Output BNC to the Video Input BNC on the LBA-PC.

All other operating requirements for the Pyrocam I are still applicable, and must be set up correctly. Don’t forget to periodically calibrate your Pyrocam.

Hint: First set up your Pyrocam to operate in one of its stand-alone configurations, and then connect it to the LBA- PC only after you’re sure that it is operating satisfactorily with your laser.

Operator’s Manual

27

LBA-PC

Doc. No. 10654-001, Rev 4.10

 

 

Page 27
Image 27
Sigma LBA-712 Special Setup for Pyrocam I Operation, Pyrocam I with Non-Digital LBA-PC’s, Pyrocam I Setup Requirements