A.2 IEEE 488.2 Common Commands

The following IEEE 488.2 common and related commands are supported by the LBA-PC.

 

Command

Meaning

Usage

 

 

 

 

 

 

 

*IDN?

Identification

LBA-PC returns the string “Spiricon, LBA-

 

 

 

 

PC, ssss, v.vv” where ssss is the serial

 

 

 

 

number of the board and v.vv is the

 

 

 

 

software version number.

 

 

 

 

 

 

 

*RST

Reset

Causes the LBA-PC to do the following:

 

 

 

 

Clear ELR, ESR, STB

 

 

 

 

Stop running

 

 

 

 

Restore last configuration saved or

 

 

 

 

restored

 

 

 

 

 

 

 

:CLR

Clear Queues

Clear output queue (MAV) and error

 

 

 

and Status

message queue (EMAV). Clear ELR, ESR,

 

 

 

Registers

STB.

 

 

 

 

 

 

 

*CLS

Clear Status

Clear ELR, ESR, STB.

 

 

 

Registers

 

 

 

 

 

 

 

 

:ELE

300PC Event

Set mask to enable event status

 

 

 

Status Enable

notification of corresponding bits set in

 

 

 

 

ELR.

 

 

 

 

 

 

 

:ELE?

 

Return ELE contents.

 

 

 

 

 

 

 

:ELR?

300PC Event

Return and clear ELR contents:

 

 

 

Status

bit 7 - Range Error flag, set when host

 

 

 

Register

 

 

 

attempts to set an input to a value that

 

 

 

 

 

 

 

 

exceeds permissible range.

 

 

 

 

bit 6 - Checksum Error flag, set when

 

 

 

 

locally computed checksum does not

 

 

 

 

match transmitted value.

 

 

 

 

bit 5 - unused

 

 

 

 

bit 4 - unused

 

 

 

 

bit 3 - unused

 

 

 

 

bit 2 - New Frame Available flag, set each

 

 

 

 

time the LBA-PC has acquired a new

 

 

 

 

frame of data.

 

 

 

 

bit 1 - Results Available flag, set each

 

 

 

 

time the LBA-PC completes a set of

 

 

 

 

computations for a new frame of data.

 

 

 

 

bit 0 - Results Pass/Fail flag, set

 

 

 

 

whenever a pass/fail alarm occurs while

 

 

 

 

running.

 

 

 

 

 

 