Sigma LBA-500PC, LBA-710, LBA-714PC, LBA-300, LBA-712, LBA-708, LBA-700, LBA-400 manual Lba-Pc

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 61
Image 61

Figure 19

Operating hints:

Increasing Video Gain also increases video noise. Use Video Gain sparingly, or not at all. Leave it set to 1 whenever possible.

The Black Level will be adjusted automatically each time you perform an Ultracal! calibration cycle. Therefore, it is best that you make no changes to this setting. If you alter the Black Level after an Ultracal!, the calibration will no longer be valid.

3.2.4.3.1The Four Camera Option

Some enhancements have been made to the four-camera option with the release of software version 4.00. These enhancements make it possible to automatically sequence data collection between multiple cameras (up to four). This sequencing only applies to analog cameras, and the following operating conditions must be met:

All cameras must be the same type or model.

The cameras must be Gen-Locked together.

If cameras with remotely controlled shutters are used the shutter control circuits must be paralleled to each camera.

The status of the input video is shown in the video enunciator located at the bottom of the LBA-PC’s main display window. The number shown in this enunciator indicates which camera is currently selected. If it shows GREEN then camera sync is present. If it shows RED then there is either no camera connected, or the connected camera is not operating.

3.2.4.3.2How to Specify Video Gain, Black Level, Shutter, and Video Trigger Level

On the right side of the Camera panel you will see a panel containing a Camera# drop- down, Video Gain, Black Level, and Shutter controls. Click on the drop-down to select which camera is affected, the current settings are displayed, and then make the desired changes. Note that Video Trigger Level, Trigger section below, can also be set for each camera.

On the toolbar, changing the Video Gain, Black Level, Shutter, and Video Trigger Level affects only the current camera listed in the video enunciator located at the bottom of the main display window.

Operator’s Manual

61

LBA-PC

Doc. No. 10654-001, Rev 4.10

 

 

Page 61
Image 61