Command Descriptions
2–96 TLS 216 Programmer Manual
NOTE. The DIAg:STATE EXECute command may take 30 seconds or longer to
respond. This command performs a warm boot and does not return control to the
instrument controller until diagnostics are complete.
Calibration and Diagnostic
 

  
 runs the diagnostic test sequences specified by the DIAg:SELect
commands. When complete, the logic scope will return to its pretest state. If the
PON event was enabled before running the tests, a Service Request will be
generated. When the Service Request has been received, the pass/fail status of
the tests can be returned by executing the DIAg:RESUlt:FLAg? query.
The DIAg:STATE EXECute command clears the following:
HEvent Queue
HInput Queue
HStatus Registers (SESR and SBR)
To enable a power-on event to generate a Service Request, send the following
commands before running diagnostics:
HDESE 128
H*ESE 128
H*SRE 32
H*PSC 0
 
executes all selected diagnostic tests.
DISplay? (Query Only)
Returns the current display settings.
Display
Group
Syntax
Arguments
Examples
Group