Command Descriptions
2–190 TLS 216 Programmer Manual
MEASUrement :REFLevel ABSolute
:?
MEASUREMENT:REFLEVEL:ABSOLUTE?
might return a string similar to: :MEASUREMENT:REFLEVEL:ABSOLUTE:HIGH
0.0E+0;LOW 0.0E+0;MID 0.0E+0;MID2 0.0E+0.
MEASUrement:REFLevel:ABSolute:HIGHSets or queries the High reference level, and is the 100% reference level when
MEASUrement:REFLevel:METHod is set to ABSolute. This command is
equivalent to setting the Reference Levels in the Measure menu.
Measurement
MEASUrement:REFLevel:ABSolute:HIGH <NR3>
MEASUrement:REFLevel:ABSolute:HIGH?
HIGH
?
<Space> <NR3>
MEASUrement :REFLevel :ABSolute :
<NR3> is the high reference level in volts. The default is 0.0 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:HIGH 1.71E+0
sets the high reference level value to 1.71 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:HIGH?
might return 4.0E+0, indicating the high reference level value is 4.0 volts.
MEASUrement:REFLevel:ABSolute:LOWSets or queries the Low reference level, and is the 0% reference level when
MEASUrement:REFLevel:METHod is set to ABSolute. This command is
equivalent to setting the Reference Levels in the Measure menu.
Measurement
Examples
Group
Syntax
Arguments
Examples
Group