Command Descriptions
2–244 TLS 216 Programmer Manual
OUTRange specifies that the trigger inputs must meet the Boolean requirements
of the Delayed Trigger pattern definition for a period of time that ends either
before or after the user specified time range. The range is 2.0 ns to 1.0 second.
TRIGGER:DELAY:PATTERN:RANGE:WHEN INRANGE specifies that the channel
inputs must cease to meet the Boolean requirements of the delayed trigger
pattern definition within the time range defined.
TRIGGER:DELAY:PATTERN:RANGE:WHEN INRANGE?
might return INRANGE, indicating the channel inputs must meet the delayed
trigger pattern requirements within the specified time range.
TRIGger:DELay:PULse? (Query Only)Returns the delayed pulse trigger parameters.
Trigger
TRIGger:DELay:PULse?
:DELayTRIGger : PULse ?
TRIGGER:DELAY:PULSE?
might return TRIGGER:DELAY:PATTERN:CLASS WIDTH;TRIG WHEN
WITHIN;SOURCE CH1;POLARITY POSITIVE as the current delayed pulse
trigger parameters.
TRIGger:DELay:PULse:CLAssSets or queries the class used for delayed pulse triggering. This command is
equivalent to selecting Glitch or Width from the Delayed Pulse Trigger Class
menu.
Trigger
TRIGger:DELay:PULse:CLAss { GLItch | WIDth }
TRIGger:DELay:PULse:CLAss?
Examples
Group
Syntax
Examples
Group
Syntax