Command Descriptions
TLS 216 Programmer Manual 2–159
*LRN ?
NOTE. The *LRN? query always returns a string including command headers,
regardless of the setting of the HEADer command. The returned string is
intended to be sent back to the logic scope as a command string. The VERBose
command can still be used normally to specify whether the returned headers
should be abbreviated.
*LRN?
a partial response might look like this:
:ACQUIRE:STATE 1;MODE SAMPLE;NUMENV 10;NUMAVG 16;
REPET 1;STOPAFTER RUNSTOP;:DIAG:LOOP:OPTION ONCE;
COUNT 1;:DIAG:STATE HALT;:HEADER 1;:VERBOSE 1;
:CURSOR:FUNCTION OFF;VBARS:UNITS SECONDS;
POSITION1 1.00E-6;POSITION2 9.00E-6;SELECT CURSOR1;.
MEASUrement? (Query Only)Returns all measurement parameters for the four displayed measurements and the
immediate measurement the logic scope provides.
Measurement
MEASUrement?
MEASUrement ?
MEASUREMENT?
might return :MEASUREMENT:MEAS1:STATE 0;TYPE PERIOD;UNITS "s";
SOURCE1 CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION
FORWARDS;:MEASUREMENT:MEAS2:STATE 0;TYPE PERIOD;UNITS "s";
SOURCE1 CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION
FORWARDS;:MEASUREMENT:MEAS3:STATE 0;TYPE PERIOD;UNITS "s";
SOURCE1 CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION
FORWARDS;:MEASUREMENT:MEAS4:STATE 0;TYPE PERIOD;UNITS "s";
SOURCE1 CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION
FORWARDS;:MEASUREMENT:IMMED:TYPE PERIOD;UNITS "s";SOURCE1
CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION FORĆ
WARDS;:MEASUREMENT:METHOD HISTOGRAM;REFLEVEL:METHOD PERCENT;
ABSOLUTE:HIGH 0.0E+0;LOW 0.0E+0;MID 0.0E+0; MID2.0.0E+0;
Examples
Group
Syntax
Examples