Command Descriptions
2–188 TLS 216 Programmer Manual
MEAS
::MEASUrement UNIts
<x> ?
<QString> returns "V" for volts, "s" for seconds, "HZ" for hertz, "VV" for
volts2, or "%" for percent.
MEASUREMENT:MEAS3:UNITS?
might return "%", indicating the units for Measurement 3 are percent.
MEASUrement:MEAS<x>:VALue? (Query Only)Returns the measurement value calculated for MEAS<x>. MEAS<x> is one of
the four measurements available (named Measurement 1 through Measurement
4), where <x> specifies 1 through 4.
NOTE. This value is a display value and will be updated every 1/3 second.
Measurement
MEASUrement:MEAS<x>:VALue?
MEASUrement : MEAS ?
<x> :VALue
<NR3>, the current value of the measurement.
MEASUREMENT:MEAS<x>:VALUE?
might return 1.20E+0, indicating the measurement value is 1.20 volts.
MEASUrement:METHodSets or queries the method used to calculate the 0 and 100% reference level. This
command is equivalent to setting the High-Low Setup in the Measure menu.
Measurement
MEASUrement:METHod { HIStogram | MINMax }
MEASUrement:METHod?
Returns
Examples
Group
Syntax
Returns
Examples
Group
Syntax