Command Descriptions
2–44 TLS 216 Programmer Manual
ACQuire:NUMEnvSets or queries the number of waveform acquisitions comprising an envelope
waveform. Using this command is equivalent to setting the Envelope count in
the Acquisition Mode side menu.
Acquisition
ACQuire:MODe
0 is the number of waveform acquisitions, from 1 to 2000. The
envelope will restart after the specified number of envelopes have been acquired
or when the ACQuire:STATE RUN command is sent.
or = 0 specifies continuous enveloping.
NOTE. If you set the acquisition system to single sequence, envelope mode, and
the number of envelopes to infinity, the logic scope will envelope a maximum of
2001 acquisitions.
specifies that an enveloped waveform will show the result of combining 10
separately acquired waveforms.
might return , indicating that acquisitions are acquired infinitely for
enveloped waveforms.
Group
Related Commands
Syntax
Arguments
Examples