Command Descriptions
TLS 216 Programmer Manual 2–131
GROUP<x>:OFFSet
Sets or queries the offset in volts, that is subtracted from the specified input
group before it is acquired. The greater the offset, the lower on the display the
waveform appears. Using this command is equivalent to setting Offset in the
Vertical menu.
Vertical
GROUP<x>:POSition
GROUP<x>:OFFSet <NR3>
GROUP<x>:OFFSet?
GROUP <x> :OFFSet
?
<Space> <NR3>
<NR3> is the desired offset in volts. The valid range for vertical offset is
15.0 Offset – 1 2.0 V, settable in 5 mV increments.
GROUP1:OFFSET 0.5E+0
lowers the group 1 displayed waveform by 0.5 volts.
GROUP1:OFFSET?
might return 500.0E-3, indicating that the current group 1 offset is 0.5 volts.
GROUP<x>:POSition
Sets or queries the vertical position of the specified group. The position value is
applied to the signal before digitization. Using this command is equivalent to
setting Position in the Vertical menu or adjusting the front panel Vertical Position
knob.
Vertical
GROUP<x>:OFFSet
GROUP<x>:POSition <NR3>
Group
Related Commands
Syntax
Arguments
Examples
Group
Related Commands
Syntax