Command Descriptions
TLS 216 Programmer Manual 2–333
WFMPre:<wfm>:PT_OffSelects or returns the trigger point within the waveform record. On input <wfm>
always defaults to the reference location specified by DATa:DESTination
regardless of what is sent.
Waveform
WFMPre:<wfm>:PT_Off <NR1>
WFMPre:<wfm>:PT_Off?
?
<Space> <NR1>
WFMPre : <wfm> :PT_Off
<wfm> = CH<x>
REF<y>
GROUP<z>
REF<y>CH<x>
GROUP works on the selected channel value.
<NR1> = 0 to the record length, and is the position of the trigger point relative to
DATa:STARt when queried.
WFMPRE:REF1:PT_OFF 100
specifies 100 as the trigger point within the waveform record.
WFMPRE:CH1:PT_OFF?
might return 250, indicating the trigger position within the waveform record.
WFMPre:<wfm>:WFIdSets or returns waveform or group information pertinent to input coupling,
volts/division, time/division, acquisition mode, and record length.
The WFMPre:<wfm>:WFId command is ignored on input.
Waveform
WFMPre:<wfm>:WFId <QString>
Group
Syntax
Mnemonics
Arguments
Examples
Group
Syntax