Command Descriptions
2–60 TLS 216 Programmer Manual
Table 2–22: Commands that Affect BUSY? Response
Operation Command
Single sequence acquisition ACQuire:STATE ON or
ACQuire:STATE RUN
(when ACQuire:STOPAfter is set to SEQuence)
Hardcopy output HARDCopy STARt
might return , indicating that the instrument is busy.
*CAL? (Query Only)Instructs the logic scope to perform an internal self-calibration and return its
calibration status.
NOTE. The self-calibration takes approximately 10 minutes to complete. No other
commands will be executed during this period.
Calibration and Diagnostic
= 0 indicates that the calibration completed without any errors detected.
0 indicates that the calibration did not complete successfully.
performs an internal self-calibration and might return to indicate that the
calibration was successful.
Examples
Group
Syntax
Returns
Examples