Command Descriptions
TLS 216 Programmer Manual 2–303
TRIGger:MAIn:SEQuence:RANge:HIGHLimit
Sets or queries the value for the upper time limit of the range for the range-quali-
fied main trigger. Using this command is equivalent to selecting Upper Limit for
main sequence triggers with range selected as the time qualifier.
Trigger
TRIGger:MAIn:SEQuence:RANge:HIGHLimit <NR3>
TRIGger:MAIn:SEQuence:RANge:HIGHLimit?
TRIGger :MAIn SEQuence :
:
HIGHLimit
RANge :
?
<Space> <NR3>
<NR3> is the upper limit of the range in seconds.
TRIGGER:MAIN:SEQUENCE:RANGE:HIGHLIMIT 1.0
sets the main sequence trigger upper limit range to 1 second.
TRIGGER:MAIN:SEQUENCE:RANGE:HIGHLIMIT?
might return 100.0E-9, indicating the main sequence trigger upper limit
range is set to 100 ns.
TRIGger:MAIn:SEQuence:RANge:LOWLimit
Sets or queries the value for the lower time limit of the range for the range-quali-
fied main trigger. Using this command is equivalent to selecting Lower Limit for
main sequence triggers with range selected as the time qualifier.
Trigger
TRIGger:MAIn:SEQuence:RANge:LOWLimit <NR3>
TRIGger:MAIn:SEQuence:RANge:LOWLimit?
Group
Syntax
Arguments
Examples
Group
Syntax