Command Descriptions
TLS 216 Programmer Manual 2–241
TRIGGER:DELAY:PATTERN:LOGIC AND?
might return OR, indicating that the Delayed Pattern Trigger Logic is set to
OR.
TRIGger:DELay:PATtern:QUAlifier
Sets or queries the Time Qualifier used to define when the logic scope triggers.
Using this command is equivalent to selecting the time qualifier Limit or Range
for the delayed pattern trigger.
Trigger
TRIGger:DELay:PATtern:QUAlifier { LIMIT | RANGE }
TRIGger:DELay:PATtern:QUAlifier?
TRIGger :DELay PATtern :
:QUAlifier
<Space>
?
RANGE
LIMIT
LIMIT defines a maximum time limit for the delayed trigger event to occur.
RANGE defines a range of time for the delayed trigger event to occur.
TRIGGER:DELAY:PATTERN:QUALIFIER RANGE
specifies that the delayed trigger event must occur either within or outside a
specified amount of time.
TRIGGER:DELAY:PATTERN:QUALIFIER RANGE?
might return LIMIT, indicating there is a maximum time limit for the delayed
trigger event.
TRIGger:DELay:PATtern:RANGE? (Query Only)
Returns the delayed pattern trigger range parameters.
Trigger
TRIGger:DELay:PATtern:RANGe?
Group
Syntax
Arguments
Examples
Group
Syntax