Command Descriptions
TLS 216 Programmer Manual 2–63
CH<x>:OFFSet? (Query Only)
Returns the offset in volts that is subtracted from the specified input channel
before it is acquired. The greater the offset, the lower on the display the
waveform appears. This query is only made available to maintain compatibility
with Tektronix TDS model instruments.
The logic scope only offsets channels when included in a user-defined groups.
See the command description GROUP<x>:OFFSet on page 2–131 of this section
to set offset for any group of input channels.
Vertical
CH<x>:OFFSet?
OFFSet
CH <x> ?
:
<NR3>
CH1:OFFSET?
might return 500.0E-3, indicating that the current channel 1 offset is
0.5 volts.
CH<x>:POSition? (Query Only)
Returns the specified channel position. The position value is applied to the signal
before digitization. This query is only made available to maintain compatibility
with Tektronix TDS model instruments.
The logic scope only positions a channel when it is included in a user-defined
group. See the command description GROUP<x>:POSition on page 2–131 of
this section to set the vertical positioning for any group of input channels.
Vertical
CH<x>:POSition?
POSition
CH <x> ?
:
<NR3>
Group
Syntax
Returns
Examples
Group
Syntax
Returns