88 Captured curves
Setting up captured curves
Models SE and DL
Setting up captured curves

Setting up Captured Curves Settings and measurement set tings

1. From the start screen, select either 1/1 octave or 1/3 octave for the
measurement type. (Repeatedly press first softkey.)
Figure 6-8: Selecting measurement type for curves
2. Next, select the measurement type, by selecting View Session (use
Up/Down Arrow to select) and press Enter . Select the appropriate
meter 1 settings (using the softkeys) and set the response time, weighting,
and measurement type. (See Figure 3-4, page 27 for an example).
NOTE: It is recommended to set Meter 1 to Slow response, Z-weighting, and
measurement to Leq.
3. Press On/Off/Esc to return to the start screen.
4. Select Setup (use up/down arrows to highlight) and press Enter .
5. Select Options menu (use up/down arrows to highlight)and press Enter .
6. Select Curves menu (use up/down arrows to highlight) and press Enter .
7. For the Mode, press Enter repeatedly to select either Survey or
Evaluation setting. (Please see explanation below).
Survey Mode: dynamically evaluates the noise and reports the
appropriate noise curve. Typically used if you wish to take multiple curve
samples to evaluate and re-evaluate noise in a room/zone.
Selecting Measurement type
Press first softkey until 1/1 or 1/3
is selected