81 Additional options
Setting up reverberation
Models SE and DL
Setting up reverberation
The Reverb Time screen is used to change or view your reverberation settings.
The steps below outline how to change the parameters. Please review Table 6-1:
Reverb time set-up optionsfor an explanation of parameters.
1. From the start screen, select Setup (by using Ar rows) & press .
2. Select Options menu (by using Arrow s) and press .
3. Select Reverb Time (by using Arro ws) and press .
4. In the Reverb Time screen to change the settings, f ollow below:
To change the Method field, press to switch between
Interruptedor Impulse(when Method is selected).
To change the Averaging field, press to switch between
Measurementor Ensemble” (when Averaging is select ed).
To change the Threshold, press . Press to toggle between
Auto” or a “dB level value. To change the dB level, press
Arrows to a specific decibel level is s elected (i.e., 80 dB) (when
Threshold is selected).
To change the Measure field, press to toggle bet ween “Auto” or
an “RT” value is selected (when Measure is selected). T o change,
press Arrows to a specific decibel level is selected (i.e., 80
dB) (when Threshold is selected).
To change the Filters field, press to switch betwee n “1/1” or
“1/3” (when Filters is selected).
Figure 6-3: Reverb Time Setup screen
Reverb Time options
a. Method field
b. Averaging field
c. Threshold field
d. Measure field
e. Filters field
a
b
c
d
e