40 Setting up auto-run
Days of week setting
Models SE and DL
NOTE: To enable Auto-run, ensure the instrument is powered off and
the date and time settings were programmed for the f uture.
10. When powered off, an Auto-Run shutdown screen will appear. (Please
see section, “Auto-Run shutdown screen.)
Days of week setting
The Days of the Week (DOW) setting is a pre-programmed weekly auto-run with
specified days, start time, duration. You can pre-program two auto-on settings
with Sunday through Saturday parameters, one specific start time, and a
specified duration for each auto-on setting. (See Figure 4-4 as an example for
settings.)
NOTE: In order for the Days of the Week mode to autorun, ensure the
instrument is in the “off” position. The meter will then automatically switch into
Auto-Run based on your setup.
Figure 4-3: Example of setting up Auto-Run (QSP-II)
A study run under the DOW mode wakes up the instr ument from a powered
down condition, runs the study, saves it, then powers the instrument down
again. The cycle repeats for the next time and da y in the setup, and for the
other Auto-on 2 if that is also activated in the mode.
Setting the Date of Week (DOW)
1. From the start screen, press Up/Down to highlight Setup and
press Enter .
2. Select Auto-Run by using Up/Down to highlight and press Enter
to select.
3. In the Auto-Run screen, repeatedly press Enter until DOW appears.
4. Press the Down arrow to select View/Set Param eters and press
Enter .
Days of Week (Auto-Run) setting
To program Auto-On 1, first select
the days of the week, then specify a
start time for your test with a
duration (to end your test).