38 Auto-Run
Date setting
Models SE and DL
Auto-Run
SoundPro SE/DL has four automatic run modes which include: Date auto-
start with duration, Day of the Week auto-start with dur ation, Timed Run
(duration only setting), and Level Triggered Runs ( or Level auto-start runs).
Setting up auto-run
All Auto-Run setups begin in the Auto-Run screen. T he Auto-Run screen
shows the current status of the Auto-Run feature. The f ollowing features are
available: Disabled, Timed Run, DOW (day of week), Date, and Level-
Triggered settings.

Date setting

A Date mode run wakes up the instrument from a po wered down condition,
runs the study, saves it, then powers down again and i s only executable
once. The parameters are selectable by specified da tes and times with up to
four programmable individual setups.
A plus sign preceding a number (i.e., +1) indicates if one of the four settings
is enabled (or activated) where as a negative sign ( i.e. -4) represents a
disabled (or deactivated) setting.
Setting up or changing auto-run with date parameter
1. From the start screen, press Up/Down to highlig ht Setup and press
Enter .
2. Select Auto-Run by using Up/Down to highlight and press Enter
to select.
3. In the Auto-Run screen, repeatedly press Enter until Date appears.
Figure 4-1: Auto-Run with Date selected
4. To select View/Set Parameters, press the Down arrow to highlight and
press Enter .
Date setting
View/Set Parameters when
selected a new menu appears