75

Back-erasing

Stopping a session

Stopping a session

You can stop the session manually or with a trigger command, but only when the instrument is in the Pause state. When the session is stopped, the Stop icon , appears in the Status Area and the RunTime clock is zeroed. For information about using triggers, see “Digital Outputs/Triggers” on page 50. Stopping a session is sometimes referred to as “closing the session.” Conversely, a session is often described as “open” when the instrument is either in a run or pause state.

Stopping and pausing

A manual session stop is a little different from a study pause. To pause a

study, , you simply press the Run/Pause key and the command is immediately executed. To stop a session, , however, you need to press and hold the Stop key.

If you don’t press and hold the stop key for the 3-second countdown period, the stop command is ignored and the session remains open. If you hold throughout the countdown, the following events occur in proper sequence.

Stop icon

Stop key

Run/Pause key

Figure 5-15: Stopping and pausing

Stopped - The session is stopped.

Transferred - The session data in memory is transferred to the memory card.

Cleared - The instrument’s memory is cleared of session data.

Zeroed - The RunTime clock is cleared to zero.

Models SE and DL

Page 88
Image 88
3M SE, DL user manual Stopping a session, Stopping and pausing