Appendix A

130

Models SE and DL

denser (200 volts must be set to “On”). Use for genera l, higher-level, free-

field measurements.

QE4150 - Free-field, Class/Type 1 (optional) ½-inch diam eter (.52"), con-

denser (200 volts must be set to “On”). Use for genera l free-field measure-

ments.

QE4170 - Pressure, Class/Type 1 (optional), 1-inch di ameter (.938"), con-

denser (200 volts must be set to “On”). Use for coup ler measurements, audi-

ometer calibration, low-frequency measurements, lo w level measurements

(due to low noise floor) or as a laboratory standard .

Characteristic
BK4936
QE7052
QE4110
QE4130
QE4150
Accuracy
Class 1
Class 2
Class 1
Class 1
Class 1
Polarization
Electret
Electret
200V
200V
200V
Diameter (inches)
½
½
¼
½
½
Response
characteristic
Free- Field Free- Field Free- Field
Free- Field Free- Field
Pressure
Frequency response
(± 2 dB)
8 Hz to 20 kHz 20 Hz to 17
kHz
20 Hz to
24.5 kHz
5 Hz to 24.5
kHz
5 Hz to 20
kHz
Sensitivity (dBV)
-28
-29
-49.1
-36
-26
Sensitivity (mV)
40
35
3.5
15.8
50
dB noise (1 kHz third
-
octave band)
0 0 - 10 0 -1
dBA noise
22
22
-
32
22
dBC noise
31
31
-
41
31
dBZ noise
35
35
-
45
35
dBF noise
40
40
-
50
40
Mic range dBA min.
(recommended)
27 27 60 37 27 23
dB Peak
142
143
167
150
143
Nominal capacitance
(pF)
12 15 6.5 18 18 60
Part number
059-523
056-317
059-413
058-659
058-489
AC/DC outputs

Connector - One 3.5 mm stereo plug connector for both AC and DC outp uts.

AC output ~ Meter 1 output, and includes the frequency response s etting for

that meter. The AC output is affected by the instrum ent’s range setting.

DC output - Meter 1 output, and includes the effects of the tim e response

and frequency weighing for the meter.