94 Audiometric background curves
Setting up audiometric background curves
Models SE and DL
Audiometric background curves
The SoundPro SE/DL series supports the following t wo types of audiometric
test room sound level curves:
Audiometric Test Room Background Sound Level Curves (OSHA)
which compares the ambient noise readings to the m aximum
permissible ambient noise levels (MPANLs) at 500 H z, 1K, 4K, 8K as
specified by the OSHA Hearing Conservation Amendm ent (1983).
Audiometric Test Room Background Sound Level Cur ves (ANSI)
which compares the ambient noise to selectable crit erions specified
in ANSI S3.1-1999 and allows for a variety of audiom etric earphones
using 1/1 and 1/3 octave band analysis.

Setting up audiometric background curves

For set-up parameters, it is recommended to set Meter 1 to Slow Response,
Z-Weighting, and measurement to Leq before you run your study.
Additionally, the measurement type should be set to 1/1 octave for OSHA
audiometric testing and should be set to 1/1 or 1/3 octave for ANSI
audiometric testing. The steps below detail the proc edures.
Setting up audiometric background curves and measurement parameters
1. From the start screen, select either 1/1 octave or 1/3 octave for the
measurement type. (Repeatedly press first sof tkey.)
Figure 6-14: 1/1 or 1/3 octave measurement type for audiometric testing
Selecting measurement type
Press first softkey until 1/1 or
1/3 is selected