50

Digital outputs/Triggering

Tracking runs

Digital outputs/Triggering

All digital output signals (also referred to as “Triggering”) appear on pins of the Auxiliary Port (“Auxiliary port” on page 7). Manage these digital outputs in the Ext-Trigger screen.

Tracking runs

The instrument can be configured to use a digital output to signal the start or end of a run.

Enabling or Disabling run/pause digital outputs/triggers

1.From the start screen, press Up/Down to highlight Setup and press Enter .

2.Select Digital-Outby using Up/Down to highlight and press Enter to select.

3.

Press

to select Output 1.

4.

Press

to enable (Run/Pse) or disable (Off) run tracking.

5.If you enabled and want to reverse the trigger logic, continue below. Otherwise, skip to step 9.

6.Press to select Logic 123, then press to enter the first data field. The second data field is for Output 1.

7.Press to select the second data field, then press to toggle the logic state. The displayed logic state is for a running study.

8.Press to exit the Logic 123 data field.

9.If finished, press to exit the screen.

OUTPUT 1

Select Run/Pause to enable

Figure 4-13: Digital-Out screen

Models SE and DL

Page 63
Image 63