80

Additional options

Reverberation options

Reverberation options

The following setup options are selectable via the Reverberation Time Options Setup menu (displayed in Figure 6-3).

Field Description

 

Explanation of field selections

 

 

 

 

Method -

 

Interrupted – (This is the default method setting.)

 

Selecting the

 

Interrupted method uses a constant sound source (i.e., dodecahedral

 

calculation process

 

speaker), waiting for the level to stabilize, and then the source is turned

 

of RT60 reverb

 

off. (The method uses random noise or pink noise to test reverb. time.)

 

time.

 

Impulse Impulse method is when the user tests the room with a

 

 

 

 

 

 

broad spectrum impulsive noise, such as a starter pistol or a balloon

 

 

 

popping.

 

 

 

 

 

Averaging –

 

Measurement – The RT60 result times for each filter band of the 1/1

 

Selecting how it will

 

Octave or 1/3 Octave modes are averaged across the studies to arrive at

 

calculate Average

 

a session average RT60 for that frequency band.

 

for Reverb-time

 

Ensemble - The decay curves for each filter band of the 1/1 Octave or

 

 

 

 

 

 

1/3 Octave modes are averaged across the studies to arrive at an

 

 

 

average session decay curve for that frequency band that is then used to

 

 

 

calculate the session RT60 for that band. This results in either eight or

 

 

 

twenty-four session ensembles, one for each frequency band in 1/1

 

 

 

Octave or 1/3 Octave modes, respectively.

 

 

 

 

Threshold -

 

Auto - (This is the default threshold setting.) The instrument will

 

Specifies the dB

 

automatically chose the level and calculate RT60. When in run mode, it

 

level for triggering

 

sets the threshold relative to the noise level in the room. It provides a

 

the start or run of

 

trigger level which is customized to each band.

 

the instrument. The

 

40 dB – 140 dB (in 1dB increments). The threshold decibel is user-

 

interpretation of this

 

 

field is dependent

 

selectable between 40 dB to 140 dB in one decibel increments.

 

on the

 

For Interrupted Noise method, level below this threshold will trigger

 

measurement

 

the start of a measurement.

 

method.

 

For Impulse method, levels above this threshold will trigger the start

 

 

 

 

 

 

of a measurement.

 

 

 

 

 

Measure -

 

Auto - (This is the default Measure setting.) The instrument will

 

The expected

 

automatically pick the best Txx value based on the expected decay (i.e.,

 

decay value

 

T20).

 

 

 

T60 (T10 – T60) – (The default setting is Auto.) The selected “Txx

 

 

 

value is the actual amount of drop/decay the instrument measured. The

 

 

 

selectable ranges are T10-T60.

 

Filters -

 

Measurements are plotted and stored in 1/1 or 1/3 octave band chart

 

1/1 or 1/3 bands

 

and decay chart. This field is only selectable if both 1/1 and 1/3 octave

 

 

 

filter bands are installed features. Otherwise, the instrument will apply

 

 

 

the installed filter.

 

Table 6-1: Reverberation time set-up options