108 Speech Intelligibility

STI-PA results screens

CAP-1

Press first softkey to assign a captured curve name (i.e., CAP- 1, CAP-2, CAP- 3, or CAP-4)

Curve indicator

Capture Press Capture softkey to save the background noise/SPL.

Figure 6-26: Captured curves/background noise

5.Once all background noise has been captured, refer to, “Applying captured curves” below.

Applying captured curves for STI-PA testing

To apply the curves during a STI-PA measurement remember to complete one and two below before applying the curve.

Applying captured curves

1.From the Start screen, select View Session and press .. Press Altf key.

2.Select one of the captured curves by pressing the 1st softkey (i.e, Cap-1). Press On/Off/Esc key.

3.From the Start screen, press the 1st softkey continuously until STI-PAis selected. Then, select View Session and the STI-PA screen will open.

POST CAP-1

The selected Captured Curve, i.e., CAP-1, will appear on the STI-PA screen. To change the Cap-x, repeat steps 1-2 above.

Figure 6-27: Post CAP-1 applied during STI-PA test

Models SE and DL

Page 121
Image 121
3M DL, SE user manual Applying captured curves for STI-PA testing,  Applying captured curves