Chapter 5: Troubleshooting

To troubleshoot attenuator failures

Attenuator Assembly

Troubleshooting

Perform Connectivity Test on all channels

Set attenuator

contact counter to zero for channels where attenuator assembly replaced

Replace faulty

attenuator assembly

and put good

assembly back in original channel.

 

 

 

 

 

 

 

Swap failing channel

 

 

 

 

 

 

 

 

attenuator assembly

 

Do channels fail?

 

Yes

 

 

 

with known good

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

assembly.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Perform Click Test on

 

 

 

 

 

all channels

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Perform Connectivity Test on all channels

Does same

No

channel fail?

Yes

 

 

 

 

 

 

 

Swap failing channel

 

 

 

 

 

 

 

 

 

 

attenuator assembly

 

Perform Click Test on

Do channels fail?

 

Yes

 

 

 

 

with known good

 

all channels

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

assembly.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

No

Set attenuator

contact counter to

zero

Run Self-Calibration

Go to Attenuator Assembly

Troubleshooting if assemblies

replaced

Does same

assembly fail?

Write attenuator

contact information on attenuator labels of defective assembly

Yes

Exchange or replace acquisition assembly

5–58

Page 135
Image 135
Agilent Technologies 45A, 46A, 54835A manual Attenuator Assembly Troubleshooting