Agilent Technologies 54835A, 45A, 46A manual To remove and replace the AutoProbe assembly

Models: 54835A 46A 45A

1 207
Download 207 pages 4.15 Kb
Page 143
Image 143
To remove and replace the AutoProbe assembly

Chapter 6: Replacing Assemblies

To remove and replace the AutoProbe assembly

To remove and replace the AutoProbe assembly

Use this procedure to remove and replace the AutoProbe assembly. When necessary, refer to other removal procedures.

1Disconnect the power cable and remove the cover.

2Remove the AutoProbe connector assembly, the subpanel, and the probe connector assembly that fits around the front-panel BNC connectors, by doing the following:

a Disconnect the mylar ribbon cable W13 from the Probe Power and Control Board, A11.

See “To disconnect and connect Mylar flex cables” in this chapter.

Figure 6-4

Disconnecting W13

bLocate the access hole in the front-panel assembly below and almost between channel two and channel three attenuators.

Figure 6-5

Access Hole

6–6

Page 143
Image 143
Agilent Technologies 54835A To remove and replace the AutoProbe assembly, Disconnect the power cable and remove the cover